Devis Casaro1, Laura Campochiaro1, Diego Copelli1, Lorena Gasparini1, Salvatore Grillo1, Marco Piedigrosso1, Tatiana Salvo1, Andrew Nicholas Scott1
1Chiesi Farmaceutici SpA, Parma, Italy
Summary
The aerodynamic particle size distribution (APSD) is a critical quality attribute (CQA) for orally inhaled and nasal drug products (OINDPs). The APSD is typically assessed in vitro using cascade impactors such as the Next Generation Impactor (NGI). The use of an NGI for APSD determination of dry powder inhalation products (DPI) is particularly challenging, since various factors can influence inhalation performance. The aim of this study was to investigate the intra- and inter-laboratory method variability and identify potential variables affecting the APSD analysis using NGI. The tested product was a lactose-based reservoir, device metered DPI. The key findings showed that the increase of humidity from 45% to 65% RH during the NGI test significantly decreased the Fine Particle Mass (FPM) and increased Mass Median Aerodynamic Diameter (MMAD), FPM decreased also with the device conditioning time from 3 to 7 days. These effects are likely driven by tribocharging phenomena, which are known to be influenced by both humidity and conditioning time. Additionally, changing the NGI cup coating from 1% to 2% Tween 20 increased MMAD. The device orientation had no effect on FPM and MMAD responses. The observed allowed us to establish the Analytical Target Profile, quantify variabilities impact as well as to set the control strategy, allowing us to meet regulatory expectations for lifecycle method management.