Particle size distribution, delivered dose and lung deposition of two mesh nebulisers guiding patients on slow and deep inhalation

Author:

Type: Poster

SUMMARY

The FOX Inhalation System is a novel, breath-actuated mesh nebuliser that guides inhalation flow rate and volume to optimise drug delivery to the lungs and provides real-time feedback to improve patients inhalation technique. In order to characterise the drug lung delivery efficiency of the FOX Inhalation System (Vectura, UK), aerosol characteristics were assessed by cascade impaction testing for Salbutamol, Cromoglicic Acid and Ipratropium bromide and regional respiratory deposition was modelled. To compare the aerosol characteristics and performance with other mesh nebulisers, which are well characterised in terms of lung delivery the AKITA² APIXNEB (Vectura, UK) was included into this study as a comparator.  Mean values for particle size (MMAD) of FOX were 3.97 to 4.04 µm and deviated from AKITA² APIXNEB by less than 14% for the tested drug products. Delivered dose results were 88.8-96.0% of filled dose with less than 3.3% deviation from AKITA² APIXNEB. Fine particle fraction deviated for FOX and AKITA² APIXNEB and was 61.5 to 63.0% for FOX and 50.9-54.6% for AKITA² APIXNEB, which was 15 to 24% higher for FOX. Fine particle dose was 3 to 21% higher for FOX. Intrathoracic deposition from deposition modelling was similar between the two devices and were 67.5 to 72.7% of filled dose.